Configuration
thermoIMAGER TIM 48
Flag:
Each pixel of the detector array emits its individual signal level depending on the ambient temperature. In
order to calculate these individual levels the detector has to be exposed to a surface with a homogenous
temperature. Therefore a shutter (or flag) is released in regular intervals.
The flag can be also used to protect the detector from potentially dangerous radiation (for example laser
radiation).
Min. interval controls that the flag will not be released before given time period even thought it might be
required by the imager. If the Max. interval is defined the flag will be always released after certain time
period even if not required by the imager.
Emissivity / Ambient temperature:
You may enter fixed emissivity/ambient temperature value.
By default the ambient temperature value is delivered by the camera’s internal sensor
In case that the values are controlled through Process Interface (PIF) the here defined values will be
ignored.
[ ► Emissivity ]
Reference temperature
There are two options available to define the reference temperature – by entering fixed value or using the
Process Interface input (PIF in).
Using comparisons of reference temperature source automatic or user-defined adjustments of object
temperature can be carried out.