21
Model CTLF-SF50-C3
Optical resolution 50:1
Temperature range
1
-50°C to 975°C
Spectral range 8 to 14µm
System accuracy
2,3
<1.5% or <1.5°C
Repeatability
2
<1% or <1°C
Temperature resolution
0.5°C
Response time (90% signal)
9ms
Emissivity/gain
1
0.100 to 1.100
Transmissivity/gain
1
0.100 to 1.000
Signal processing
1
peak hold, valley hold, average; extended hold function with threshold and hysteresis
Certificate of calibration optional
Outputs/analog
channel 1 0/4 to 20mA, 0 to 5/10V, thermocouple J, K
channel 2 sensor temperature (-40 to 85°C as 0 to 5V or 0 to 10V), alarm output
optional relay: 2 x 60VDC/ 42VACeff; 0.4A; optically isolated
Alarm output
open - collector (24V/ 50mA)
Outputs/digital optional
USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output mA max. 500Ω (with 5 to 36VDC)
voltage output
mV min. 100kΩ load impedance; thermocouple 20Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger (reset of hold functions)
Cable length
3m (standard), 8m, 15m
Power supply
8 to 36VDC; max. 160mA
Laser class II (635nm), 1mW, ON/OFF via controller or software
Environmental rating IP 65 (NEMA-4)
Ambient temperature
sensor: -20°C to 85°C (50°C if Laser ON) controller: 0°C to 85°C
Storage temperature
sensor: -40°C to 85°C controller: -40°C to 85°C
Relative humidity
10 to 95%, non condensing
Vibration sensor
IEC 68-2-6: 3 G, 11 to 200Hz, any axis
Shock sensor
IEC 68-2-27: 50 G, 11ms, any axis
Weight
sensor: 600g; controller: 420g
1
adjustable via controller or software
2
± ambient temperature: 23 ±5°C; whichever is greater
3
temperature of the object >0°C
Accessories page 30 - 31
Mounting bracket
Air purge collar
Rail mount adapter for controller
Water cooled housing
Interface kit
Software CompactConnect
Certificate of calibration
Cable length [3 m Standard / 8 m / 15 m]
Focus [SF50 / CF1 / CF2 / CF3 / CF4]
thermoMETER CTLaserFAST
Product identification
CTLF - SF50 - C3