37
Model CTG-SF10-C3 CTG-SF20-C3
Optical resolution 10:1 20:1
Temperature range
1
100 to 1200°C 250 to 1650°C
Spectral range
5.2µm
System accuracy
2
<1% or <2°C
Repeatability
2
<0.5% or <0.5°C
Temperature resolution
<0.1°C <0.2°C
Response time
80ms 120ms
Emissivity/gain
1
0.100 to 1.100
Transmissivity/gain
1
0.100 to 1.100
Signal processing
1
Peak hold, valley hold, average; extended hold function with threshold and hysteresis
Outputs/analog
channel 1 0/4 to 20mA, 0 to 5/10V, thermocouple J, K
channel 2 sensor temperature (-40 to 85°C as 0 to 5V or 0 to 10V), alarm output
optional relay: 2 x 60V DC/ 42VAC
eff
; 0.4A; optically isolated
Outputs/digital
optional
USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output
mA max. 500Ω (8 to 36VDC)
voltage output mV min. 100kΩ load impedance
thermocouple 20Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger
Cable length 3m (standard), 8m, 15m
Power supply 8 to 36VDC; max. 100mA
Environmental rating IP 65 (NEMA-4)
Ambient temperature sensor: -20°C to 85°C controller: 0°C to 85°C
Storage temperature sensor: -40°C to 85°C controller: -40°C to 85°C
Relative humidity 10 to 95%, non condensing
Vibration
sensor
IEC 68-2-6: 3 G, 11 to 200Hz, any axis
Shock
sensor
IEC 68-2-27: 50 G, 11ms, any axis
Weight sensor: 42g; controller: 420g
1
adjustable via programming keys or software
2
± ambient temperature: 23±5°C; whichever is greater
Accessories page 48 - 51
CF lense
Protective window
Mounting bracket / Mounting bolt
Air purge collar
Right angle mirror
Rail mount adapter for controller
Massive housing
Protective tube
Laser sighting tool
Digital-Interface kit
Software CompactConnect
Certificate of calibration
Relay output module
Product identification
CTG - SF10 - C3