27
Cable length [3 m Standard / 8 m / 15 m]
Focus [SF / CF1 / CF2 / CF3 / CF4]
Temperature range [L / H]
Spectral range [2,3 µm]
thermoMETER CTLaserM3
Product identification
CTLM - 3 L SF60 - C3
Model
CTLM-
3LSF60-C3
CTLM-
3HSF100-C3
CTLM-
3H1SF300-C3
CTLM-
3H2SF300-C3
CTLM-
3H3SF300-C3
Optical resolution 60:1 100:1 300:1 300:1 300:1
Temperature range
1,2
50 to 400°C 100 to 600°C 150 to 900°C 200 to 1200°C 350 to 1800°C
Spectral range 2.3 µm
System accuracy
3
<(0.3% of reading +2°C)
Repeatability
3
<(0.1% of reading +1°C)
Temperature resolution (digital) 0.1°C
Response time (90% signal)
4
1 ms
Emissivity/gain
1
0.100 to 1.100
Transmissivity/gain
1
0.100 to 1.100
Signal processing
1
peak hold, valley hold, average; extended hold function with threshold and hysteresis
Certificate of calibration optional
Outputs/analog
channel 1
0/4 to 20mA, 0 to 5/ 10V, thermocouple J, K
sensor temperature (-40 to 85°C as 0 to 5 V or 0 to 10 V), alarm output
channel 2
Outputs/analog (option) relay: 2 x 60 VDC / 42 VAC; 0.4 A; optically isolated
Alarm output open-collector (24 V / 50 mA)
Outputs/digital option USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output mA max. 500Ω (with 5 - 36VDC)
voltage output mV min. 100kΩ load impedance; thermocouple 20Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger (reset of hold functions)
Cable length 3 m (standard), 8 m, 15 m
Power supply 8 to 36 VDC; max. 160 mA
Laser class II (635nm), 1mW, ON/OFF via controller or software
Environmental rating IP 65 (NEMA-4)
Ambient temperature sensor: -20°C to 85°C (50 °C if laser ON) controller: 0 °C to 85 °C
Storage temperature sensor: -40 °C to 85 °C controller: -40 °C to 85 °C
Relative humidity 10 to 95%, non condensing
Vibration sensor IEC 68-2-6: 3 G, 11-200Hz, any axis
Shock sensor IEC 68-2-27: 50 G, 11 ms, any axis
Weight sensor: 600 g; controller: 420 g
1
adjustable via controller or software
2
target temperature > sensor temperature + 25°C
3
E=1, response time 1s; ± ambient temperature: 23 ±5°C
4
with dynamic adaptation at low signal levels
Accessories page 26 - 27
Mounting bracket
Air purge collar
Rail mount adapter
for controller
Water cooled housing
Interface kit
Software CompactConnect
Certificate of calibration
Optical specifications thermoMETER CTlaserM3
= smallest spot size (mm)
Close Fokus
3LCF4 60:1
20 18.7 17.3 15.9 14.5 13.1 11.7 10.3 9 7.5 10.6 17 23 29
3HCF4 100:1
20 18.3 16.6 14.9 13.2 11.4 9.7 8 6.3 4.5 7.3 13 19 24
3 H1/H2/H3 CF4 300:1
20 18 16 13.8 11.8 9.7 7.6 5.6 3.5 1.4 3.8 8.6 13.3 18
distance (mm) 0 50 100 150 200 250 300 350 400 450 500 600 700 800
Far Field
3 H1/H2/H3 FF 300:1
20 19 18 17 16 15 14 13.4 12 16.5 24.4 33.4 40
distance (mm) 0 450 900 1350 1800 2250 2700 3000 3600 4000 5000 6000 6750