Model CTLG-SF45L-C3 CTLG-SF70H-C3
Optical resolution 45:1 70:1
Temperature range
1
100 to 1200°C 250 to 1650°C
Spectral range
5,2µm
System accuracy
2
<1% or <1°C
Repeatability
2
< 0.5% or <0.5°C
Temperature resolution
0.1°C 0.2°C
Response time (90% signal)
120ms 80ms
Emissivity/gain
1
0.100 to 1.100
Transmissivity/gain
1
0.100 to 1.000
Signal processing
1
peak hold, valley hold, average; extended hold function with threshold and hysteresis
Certificate of calibration
optional
Outputs/analog
channel 1 0/4 to 20mA, 0 to 5/10V, thermocouple J, K
channel 2 sensor temperature (-40 to 85°C as 0 to 5V or 0 to 10V), alarm output
optional relay: 2 x 60VDC/ 42VACeff; 0.4A; optically isolated
Alarm output
open - collector (24V/ 50mA)
Outputs/digital optional
USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output mA max. 500Ω (bei 5 to 36VDC)
voltage output
mV min. 100kΩ load impedance; thermocouple 20Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger (reset of hold functions)
Cable length
3m (standard), 8m, 15m
Power supply
8 to 36VDC; max. 160mA
Laser class II (635nm), 1mW, ON/OFF via controller or software
Environmental rating IP 65 (NEMA-4)
Ambient temperature sensor: -20°C to 85°C (50°C if Laser is ON) controller: 0°C to 85°C
Storage temperature sensor: -40°C to 85°C controller: -40°C to 85°C
Relative humidity 10 to 95%, non condensing
Vibration
sensor
IEC 68-2-6: 3 G, 11 to 200Hz, any axis
Shock
sensor
IEC 68-2-27: 50 G, 11 ms, any axis
Weight sensor: 600g; controller: 420g
1
adjustable via controller or software
2
± ambient temperature: 23 ±5°C; whichever is greater
Accessories page 30 - 31
Mounting bracket
Air purge collar
Rail mount adapter for controller
Water cooled housing
Interface kit
Software CompactConnect
Certificate of calibration
Cable length [3 m Standard / 8 m / 15 m]
Focus [SF45L/ SF70H / CF1L/H / CF2L/H / CF3L/H / CF4L/H]
thermoMETER CTLaserGLASS
Product identification
CTLG - SF45L - C3
70
89
30
100
M48x1.5
Vista de pagina 22