Vista de pagina 24
25
Model
CTLM-1LSF150-C3 CTLM-1HSF300-C3 CTLM-2LSF150-C3 CTLM-2HSF300-C3
Optical resolution 150:1 300:1 150:1 300:1
Temperature range
1
485 to 1050°C 650 to 1800°C 250 to 800°C 385 to 1600°C
Spectral range 1µm 1.6µm
System accuracy
2
<(0.3% of reading +2°C)
Repeatability
2
<(0.1% of reading +1°C)
Temperature resolution 0.1°C 0.2°C 0.1°C 0.2°C
Response time (90% signal)
3
1ms
Emissivity/gain
1
0.100 to 1.100
Transmissivity/gain
1
0.100 to 1.000
Signal processing
1
Peak hold, valley hold, average; extended hold function with threshold
and hysteresis
Certificate of calibration
optional
Outputs/analog
channel 1 0/4 to 20mA, 0 to 5/ 10V, thermocouple J, K
channel 2 sensor temperature (-40 to 85°C as 0 to 5V or 0 to 10V), alarm output
optional relay: 2 x 60 VDC/ 42VAC
eff
; 0.4A; optically isolated
Alarm output open-collector (24V/ 50mA)
Outputs/digital optional USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output mA max. 500Ω (with 5 - 36VDC)
voltage output mV min. 100kΩ load impedance
thermocouple 20Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger (reset of hold functions)
Cable length
3m (standard), 8m, 15m
Power supply
8 to 36VDC; max. 160mA
Laser class II (635nm), 1mW, ON/OFF via controller or software
Environmental rating IP 65 (NEMA-4)
Ambient temperature
sensor: -20°C to 85°C (50 °C if laser ON) controller: 0°C to 65°C
Storage temperature
sensor: -40°C to 85°C controller: -40°C to 85°C
Relative humidity
10 to 95%, non condensing
Vibration sensor
IEC 68-2-6: 3 G, 11-200Hz, any axis
Shock sensor
IEC 68-2-27: 50 G, 11ms, any axis
Weight
sensor: 600g; controller: 420g
1
adjustable via controller or software
2
E=1, response time 1s; ± ambient temperature: 23 ±5°C
3
with dynamic adaptation at low signal levels
Accessories page 30 - 31
Mounting bracket
Air purge collar
Rail mount adapter for controller
Water cooled housing
Interface kit
Software CompactConnect
Certificate of calibration
Software CompactConnect
- Display, graphic charting and recording of temperature readings
- Easy system configuration and sensor calibration
- Sophisticated signal processing features
- Programming of input and output channels
Cable length [3 m Standard / 8 m / 15 m]
Focus [SF / CF2 / CF3 / CF4 / FF]
Temperature range [L / H]
Spectral range [1 µm / 1,6 µm]
thermoMETER CTLaserM
Product identification
CTLM - 1 L SF150 - C3
70
89
30
100
M48x1.5